Model 20189/340W - Design Specifications

Isolated Probe System

340W SPECIFICATIONS
Reference conditions: zero-based spans, ambient temperature 28ºC (82ºF), D/A trim values equal to span end points, silicon fill, 316L SS diaphragm, one-second damping.

Accuracy

  • ±0.25% of calibrated span for spans from 1:1 to 10:1 of URL
  • ±(0.178+0.0072[URL/span])% of calibrated span for spans from 10:1 to 45:1 of URL
Note: Accuracy includes the combined effects of linearity, repeatability, and hysteresis

Ambient Temperature Effect

  • ±0.25% of URL per 28ºC (82ºF) shift

Temperature Limits

Sensor Assembly

  • -40 to 257ºF (-40 to 125ºC)
Electronics
  • 0 to 130ºF (-18 to 54ºC)

Static Pressure Effect

Zero

  • Eliminated at operating pressure
Span
  • Correctable to 0.2% per 150 psi (1034 kPa)

Maximum Working Pressure

  • ±150 psi (1034 kPa)
Note: MWP is the maximum pressure (static or otherwise) that can be applied to the cell without damage.

Stability

Zero

  • 0.1% of URL for 12 months
Span
  • No span drift

Humidity

  • 0-100% relative humidity, noncondensing

Vibration Effect

  • Less than ±0.05% of URL per G for 0-60 Hz in any axis up to 2 Gs maximum

Power Supply Effect

  • Less than ±0.005% of output span per volt

EMI/RFI Susceptibility

  • Less than ±0.25% of URL at 30 V/m, 30MHz-1Gz

FUNCTIONAL SPECIFICATIONS
Ranger Sensor Limits

Range Code
 D
__Minimum Span__
10 in H20(2.49 kPa)

______LRL______
-450 in H2O(-112 kPa)
______URL______
450 in H2O(112 kPa)

Zero Elevation & Suppression

  • The range may be set anywhere between the LRL and URL of the transmitter, as long as the calibrated span does not exceed the minimum allowable span. Zero and soan in the XTC are noninteractive.

Electronic Damping

  • Adjustable between 0 and 30 seconds

Transmitter Outputs

  • Analog, two-wire mA
  • Digital, HART Communications
  • Optional Transient Suppressor

Power Supply Requirements

  • Minimum Terminal to Terminal Compliance Voltage - +10Vdc
  • Maximum Terminal to Terminal Voltage - +42Vdc
  • Maximum Load - R=50º(supply voltage) - 500 ohms
Note: To ensure digital communication, HART imposes the following restrictions on the loop: 250< R < 1100 ohms, ripple < 0.2 Vp-p ac between 47 and 125 Hz, noise < 0.6 mVrms.

Local Indication

  • Optional 4-1/2 digit Smart Display

Surge Protection (with optional transient suppressor)

  • Maximum Clamping Voltage (either terminal to case)
  • DC - 68V
  • 100kVuS surge - 70 Vp
  • 1000 kVu surge - 120 Vp

Recommended Wire Type

  • <5000 ft(1524m): Belden 8641, 24AWG
  • >5000 ft(1524m): Belden 8762, 20AWG

MECHANICAL SPECIFICATIONS
Dimensions

  • See Figure 1-1

Weight(approximate)

  • 8 lbs(3.6 kg)

Electronics Housing

  • Epoxy Powder Coated, Low Copper Cast Aluminum NEMA 4X, IP66/68(submersible up to 3 meters, indefinitely)
  • (2) 1/2º - 14 NPTF Electrical Conduit Entrances (M20 x 1.5 optional)

Process Wetted Parts

  • Body: 316 SS
  • Diaphragm: 316L SS

Process Connection

  • (2) 0.75" diameter by 10.5" pressure sensors (14.5" optional)

Back to Isolated Probe System


Copyright © 1997 BIF
Comments regarding this site, please contact webmaster@intersoftgroup.com
Site created by Jon'taar Graphxs